img

Trace Oxygen Analyzer

We provide the good solution for your specific application

img

Description :


Trace Oxygen Analyzer BI 310 measures Traces of Oxygen in percent level in process gas. For O2 measurement Electrochemical technology is used.

Easy to maintain and service from user point of view. LCD, LED & Touchscreen display available. Analog 4-20 mA output with 2 nos. potential free relay contact. It has 4 different mounting configurations, Wall, Panel, Rack & Tabletop.

SPECIFICATIONS :


Method of Detection Electrochemical sensor for O2
Range 0-30% scalable
Resolution 0.01%
Accuracy & Repeatability ± 1% of FS
Drift <2% of full scale per month
Response Time (T-90) ≤30 seconds
Digital Interface RS485, MODBUS Protocol
Power 80 to 230 VAC 50 Hz

Features:-

  • 32-Bit Cortex-M3 ARM Processor
  • Long life Electrochemical O2 sensor
  • RS 485, MODBUS protocol
  • PTFE tubing and corrosion-resistant components for internal sample health
  • Flow meter for flow regulation
Top